Centrifuge module interface
Table of Contents

Advertisement

System Operation
8. Press the robot computer power switch and wait approximately
2 minutes for the robot computer to power up.
Figure 5:
1. Robot computer power switch
9. Close the panels.
10. On the LineMaster workspace station buttons, select Centrifuge.
The Details screen is displayed. The Centrifuge module is in a
Shutdown state.
11. On the States panel, select Online.
page 38
Centrifuge Robot Computer Power Button
The interface transitions to the Online state.
The automation software and the Centrifuge module software
check communication links and safety features.
The robot surveys the Centrifuge module to check for sample tubes
in the Centrifuge.