Specifications
•
Forces up to 4000 RCF
•
Centrifuging times between 1 and 999 minutes
Sample Tube Specifications
The following table lists supported tubes for the Centrifuge module.
WARNING
Do not use glass tubes in the centrifuge module. Glass tubes can break,
causing physical harm to the operator and equipment. The Centrifuge
module supports only valid plastic sample tubes. See Appendix C,
Specifications.
Note The Decapper module does not support black and dark blue caps.
Manually remove black and dark blue caps and place sample tubes in
uncapped sections of trays on the automation system.
Table 5:
Manufacturer
Product
Name
BD
Vacutainer
BD
Vacutainer
BD
Vacutainer
BD
Vacutainer
BD
Vacutainer
Greiner
Vacuette
Greiner
Vacuette
Greiner
Vacuette
Greiner
Vacuette
Terumo
VenoSafe
Terumo
VenoSafe
Sarstedt
S-
Monovette
page 92
Sample Tubes Allowen on the Centrifuge and Decapper
Tube Size Bottom
Shape
13 x 75
Round
13 x 75
Round
13 x 100 Round
16 x 100 Round
16 x 100 Round
13 x 75
Round
13 x 75
Round
13 x 100 Round
16 x 100 Round
16 x 100 Round
13 x 100 Round
13 x 90
Round
Centrifuge Module Interface Operating Instructions
Tube
Cap Type
Material
Plastic
Conventional rubber
stopper
Plastic
Hemogard
Plastic
Hemogard
Plastic
Conventional rubber
stopper
Plastic
Hemogard
Plastic
Hemogard-like, pull
Plastic
Hemogard-like,
screw
Plastic
Hemogard-like, pull
Plastic
Hemogard-like, pull
Plastic
Hemogard-like
Plastic
Hemogard-like
Plastic
Screw
Exceptions
Green/Black
caps are not
supported
092D0001-02 Rev. A