By Sample Type Tab - Siemens ADVIA LabCell Operating Instructions Manual

Centrifuge module interface
Table of Contents

Advertisement

Centrifuge Module Interface Operating Instructions
To assign a sample tube with a specific LIS code to be centrifuged before it
proceeds to other tests, perform the following steps:
1. On the LineMaster workspace, select System > Pre and Post Tests
(Shift F5).
The By LIS Code Tab is displayed by default.
2. On the Test List area on the right, select the test ID of the sample that
must be centrifuged prior to other testing.
3. On the Pre and Post Test area on the left, on the Selected column,
select the Centrifuge pre-test check box.
4. To verify that LIS codes are added, select the Usage tab.
The information you selected is listed in the Usage tab.

By Sample Type Tab

The By Sample Type tab of the Pre and Post Tests (Shift F5) screen
assigns the Centrifuge pre-test to selected sample types.
Note For this option, the LIS order must specify sample types.
Figure 7:
092D0001-02 Rev. A
Pre and Post Tests (Shift F5)–By Sample Type Tab
System Operation
page 43

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia workcell cdx

Table of Contents