Explanation of the System Data Fields
TABLE 14.
Setting
Function/Description
GN #/#/#
Gain Use the Main knob to adjust the gain. The setting is in
this order:
5/ /13 shows a
of 13.
For
For Color Gain the range is: 1 through 30.
For Doppler Gain the range is: 1 through 42.
##fps
Frames Per Second The speed at which the frames are
displayed. Use the Angle option on the
menu to adjust the rate in 2D mode. Use the
Soft Window knob to adjust the rate in color mode.
##.#cm
2D Depth (2D, Color, and M-Mode only) Use the
Soft Window knob to adjust the depth.
The maximum scanning depth of the system is 26.9 cm.
HPF ###Hz
High Pass Filter (Doppler Mode only) Use the
Window knob to adjust the filter.
The range is from 200 Hz through 1400 Hz for
from 25 Hz through 1400 Hz for
F#
Filter (Color mode only) indicates the velocity filter applied to
the Color Doppler.
The range is F1 through F6.
Harmonic
Harmonic If your transducer has harmonic mode, the word
"Harmonic" will appear. Use the
knob to turn harmonic on or off.
The mode is
Freq#
Frequency The frequency of some transducers can be
adjusted. If your transducer can be adjusted the word "Fre-
quency" will be displayed. Use the
knob to adjust the frequency.
Invert
Invert Displayed when the spectral doppler flow display has
been inverted. The spectrum is inverted using the corre-
sponding Soft Window knob.
ACUSON Cypress™ Operator's Manual
/Color/Doppler. For example a setting of: GN
2D
gain of 5; no color gain; a Doppler gain
2D
Gain the range is: 1 through 28.
2D
when the word Harmonic is displayed.
ON
Overview of the Scanning Process
image
2D
TOOLS
FRAME RATE

DEPTH

FILTER
CW
.
PW
Soft Window

FREQUENCY

Soft Window
FREQUENCY
Soft
and
5-11