Connecting an External Monitor

WARNING
Do not use a peripheral device, such as a monitor,
within 1.5 meters of a patient unless the device receives power
from an isolation transformer that meets medical safety stan-
dards.
The Cypress system can be connected to a VGA-compatible monitor. The 15-
pin high-density "D" type connector on the I/O panel connects directly to a VGA
monitor through a standard VGA cable.
To connect an external monitor to the Cypress system, follow these steps:
Make sure Cypress system and monitor are turned
1.
Connect the monitor cable to the 15-pin connector on the Cypress system
2.
(see Figure 54).
Connect the power cord on the monitor and plug it in.
3.
Turn the Cypress system
4.
On the Cypress system, press the
5.
tab.
Use the Video Out Mode drop-down menu to select VGA.
6.
Press
to return to
SETUP
7.
Connecting an External Monitor
FIGURE 54.
ACUSON Cypress™ Operator's Manual
and then turn the monitor
ON
function key to display the Setup
SETUP
mode or press
2D
2D
15-PIN VIDEO CONNECTOR FOR
EXTERNAL MONITOR
Connecting an External Monitor
or printer
VCR
.
OFF
.
ON
to begin imaging.