Dicom Setup For Storage Commitment - Siemens ACUSON Cypress Operator's Manual

Table of Contents

Advertisement

Transferring Studies
9-12

DICOM Setup For Storage Commitment

If the optional networking package is installed, the
Server information fields must be completely defined before the ability to per-
form Storage Commitment is enabled.
To specify your Storage Commitment settings, follow these steps:
Press the
function key.
SETUP
1.
Double-click the
DICOM
2.
detailed in steps 3-9:
Click the Storage Commitment subtitle.
3.
Click the [
] button.
ADD
4.
Identify your server by typing a user-defined name in the Server Name
5.
Field.
Set the Application Entity. This defines the server application that receives
6.
Storage Commitment messages from the Cypress system.
Set the IP Address. This is the IP address of the Storage Commitment
7.
where the query will be made.
Set the Port Number. This is the remote port on the server that is to receive
8.
the connection request. The default setting is 104.
The Test Result field lets you know the test result of the connection to the
9.
Storage Commitment server. This field is used in conjunction with the [
and [
] buttons. Before using these buttons, verify with your IT depart-
VERIFY
ment that your Storage Commitment server will respond to a ping and a
Verify. Some systems will not respond to a ping due to network secu-
DICOM
rity issues. Some Storage Commitment servers may not respond to a
Verify command — consult the server's
After changing or setting up the
by clicking the [
PING
cate its results in this field. If the ping is successful, test the
tion by clicking the [
ACUSON Cypress™ Operator's Manual
directory listing and complete the Server Information
DICOM
server settings, test the connection
DICOM
] button. The system will test the connection and indi-
] button.
VERIFY
Storage Commitment
DICOM
Conformance Statement.
connec-
DICOM
]
PING
DICOM