Trigger Modes

Triggered acquisition modes may be used if the advanced triggering option has
been installed on the system. Triggered modes allow for
techniques. See Table 16 on page 5-23 for more information. The trigger mode
information is displayed beneath the System Data. The trigger controls are
assigned to the
APEX INVERT
display the triggering options and then turn the knob to adjust the settings. Note
that the Advanced Triggering option must be installed for this Soft Window to
have multiple functions.
Pwr M
3V2c
MI 0.9
TIs 0.1
Comp 2
PProc 3
GN 5/ /
45fps
16.2cm
Harmonic
1 Beat
40.0ms
3 Frames
The
Trigger Modes
FIGURE 23.
Table 15 describes the trigger modes.
Explanation of the Trigger Mode Fields
TABLE 15.
Setting
Function/Description
# Beat(s)
Beat Shows the number of beats selected for intermittent
imaging mode.
The range is from 1 through 16.
#.# ms
Time Increases the trigger time in milliseconds.
The range is from 8.0 ms through 1992.0 ms.
# Frames
Frames Shows the number of frames that are produced in
intermittent mode
The range is from 1 through 16 frames.
ACUSON Cypress™ Operator's Manual
Soft Window. Press the corresponding knob to
THE TRIGGER MODES ARE:
BEAT, TIME, AND FRAME RATE
Overview of the Scanning Process
-gated scanning
ECG
Loop 1/5
04:05:23 PM
13 June 2005
5-13