MASS STORAGE UNIT SPECIFIER DESCRIPTION
The following
describes the
use of the
mass storage unit specifier
in
the test
software.
Although
Msus$
is
referred
to specifically,
the description regarding the usage of the mass
storage unit specifier applies
to
all
three mass storage unit specifiers
Msus$
is
the
mass storage
unit
specifier which defines the
address for
the test software
When
using an
HP 9133H set
to one
volume and HP-IB address 700 or an HP 9122D set
to
HP-
IB
address 700,
Msus$
would be set as follows:
Msus$=":,700,0"..
This designates drive
0
as
the storage location for the test software-
When
using other disc drives or
another address,
you would need to
set
Msus$
to reflect the
address
of you
disc
drive-
Two
examples could
be;
Msus$=":
REMOTE 21,3"
(when
using
a
shared resource management system)
or
Msus$=":,700,0,0"
(when
using an HP
9133H set
to
more than one volume).
PERFORMING THE
MODIFICATION
After you have determined the modifications which need
to be
made
to
the system variables,
you should
perform
the following.
1
Bootup HP Basic
3 0
or 4.0 with the
appropriate
binary
extensions
needed to operate
your system
If
using
an HP 9133H or 9122D,
the
binary extensions
listed
in
paragraph 2.3
are
all
you
need
2.
LOAD"FS_MANAGER'
3
EDIT
Mass_stg_init.
4.
Scroll to the system variables which
need
to be modified and
make
the
appropriate
changes.
Enter each
line after
modifying
5-
Check the system variables
to
ensure that the modifications
you
made are correct-
6.
If
the storage
media
you
are
using for the
test software
is
a
double sided
3.5"
microfloppy,
you
will
need to PURGE
"FS_MANAGER"
7.
RE-STORE "FS_MANAGER"
HP 8340A/B
and
HP 8341A/B
How
to Configure the Software