4-13.
PULSE MODULATION ACCURACY TEST
NOTE
THIS TEST
IS
NOT
APPLICABLE
TO INSTRUMENTS
EQUIPPED WITH
OPTION 006, DELETE PULSE MODULATION.
Specification
Table 4-26.
Pulse Modulation
Accuracy Test
Specifications
For
CW mode
and frequencies
>400
MHz
only:
Minimum Internally Leveled RF Pulse Width
(Tl, RFI): 100
nanoseconds
Pulse Repetition Frequency:
Internally leveled: 100 Hz to
5
MHz
Accuracy of Internally Leveled RF Pulse
VI,
PI
(relative to
CW mode
level):
NOTE
ALC attempts
to hold pulse
amplitude to same level as leveled CW signal.
Bands
and Approximate Frequency Ranges
(GHz)
(see Frequency Ranges and Bandswitch Points for complete description)
Pulse Width
100 to
<200ns
200
to
<500
ns
s.500
ns
BandO
0.4 to
<2.3
+3/-0.3dB2
+1.5/-0.3dB2
±0.3dB
Bands
1
4i
2.3 to 26.5
+1.5/-0.3dB2
±0-3dB
±0-3dB
1
Band
4
(>20
0
to 26
5
GHz)
only applies
to
the HP
8340B
2
+15
to
+55-C-
Duty
Cycle must be
>0
01%
Description
The DUT's RF output
frequency
is
down converted to 50
MHz
using
a
mixer and
a
local oscillator.
The
50 MHz
IF
signal
is
amplified and applied
to an
oscilloscope. The
DUT
is
pulsed using
a
pulse
generator. The
IF
modulation envelope
is
positioned to
convenient horizontal graticule
lines.
The
pulse
is
turned
OFF
and the
DUT's output power
level
is
adjusted so that the
IF
carrier
is
at the
horizontal graticule
lines
established when
the
DUT was
being pulsed. The change
in
DUT
output
power
level is
the accuracy error
of
the leveled RF pulse.
HP 8340B/41B