5-15.
ALC ADJUSTMENTS (Cont'd)
Table 5-8.
High Band
ALC
Adjustment
OUT Power Level
-20
dBm
-10dBm
OdBm
10dBm
18
dBm
Access Calibration Constant Number 47 by
entering the following
(SHIFT]
[GHz]
[4] [7]
[Hi]
[SHIFT]
[MHz] [1] [2] (Hz]
[SHIFT]
[kHz] [2]
[2]
[Hz] Use [STEP]
key
to
change Cal Constant Number
to
44
Associated Adjustment
A25R34 (H-20)
Calibration Constant
#47*
Calibration Constant
#44*
A25R39(H+10)
A25R24
(+20)
Check Limit
0.3 dB
0.1
dB
0.1
dB
0.1
dB
0.2 dB
Adjustment Limit
0.2
dB
0.02 dB
0.02 dB
0.02 dB
0.05 dB
17.
If all
power meter indications are
within
the test
limits,
check at
power
levels from —20 dBm to
+20
dBm
in 5
dB steps.
18.
Press
[CW]
and enter
[1] [.] [5] [GHz].
Press
[SHIFT] [PWR SWP]
to
decouple the RF attenuator
and ALC. Enter
[0] [dBm].
Press
[SHIFT] [METER]
and
enter[-]
[1] [0] [0] [dBm]. Dashes
should be displayed
in
the
POWER
dBm display.
If
not,
rotate the RPG counter-clockwise
until
the
dashes appear
(note that the UNLEVELED annunciator
will
be on at this time).
Zero
the power
meter
in
"WATT" mode
as described
in
step
15.
The
low band (Band
0) contains broadband noise.
If
the
power
meter
is
properly zeroed,
the broadband noise
will
be
compensated
and
power
measurements
will
be more accurate. Press [1NT] (UNLEVELED annunciator off) and then
[SHIFT] [PWR SWP]
(attenuator and
ALC
should
both indicate
zero).
19.
Refer to Table 5-9
below, enter
the
appropriate
DUT power
level
(ALC
level) and
check the power
meter
indication.
Check the
indication at each
power
level and
note
the
amount
of error before
making any adjustments.
If
necessary adjust
the appropriate controls shown
in
the
table
to
adjust
the
power
level to be within
the test
limits.
Adjustments are
iterative.
Adjust
in
the following order:
0
dBm,
—10
dBm, —20
dBm, —10 dBm,
0
dBm, and
+10
dBm.
When
selecting
a
new power
level, press
[SHIFT] [PWR SWP] [desired power level] [dBm].
Table 5-9.
Low
Band ALC Adjustment
DUT Power Level
-20
dBm
-10dBm
OdBm
10dBm
There
is
no low band
+20
dBm
adjustment.
Access Calibration Constant Number
46
by entering the following [SHIFT]
[GHz]
[4] [6]
[Hz]
[SHIFT]
[MHz] [(]
[2] [Hz] [SHIFT]
[kHz]
[2] [2] [Hz]
Associated Adjustment
A25R33 (L-20)
Calibration Constant
#46*
A25R108(LOFS)
A25R38(L+10)
Check Limit
0.3 dB
0.1
dB
0.1
dB
0.1
dB
Adjustment Limit
0.2
dB
0.02 dB
0.02
dB
0.02
dB
HP 8340B/41B