HP 8340B Manual page 197

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

NOTE:
Instrument calibration constants
tell
the program whether an
attenuator
is
installed.
If
an
attenuator was
installed after time of
manufacture,
calibration constant #59 should have been
changed
to reflect the addition.
If
this cal
constant was
not
updated,
the
program
tells you that no
attenuator exists, though one
is
actually installed. To update
the cal
constants
to reflect the correct
option configuration, return to the test menu and select the [Utility] softkey. Refer to
utility
8-20
and
select the
[Chg Option] softkey.
Problem Solver
If
you
run
into difficulties during this test (due
to
an incorrect
test setup,
etc.), and the softkey menu
gives you no
ABORT
selection, press [Carry On] or other continue command
until
you
reach
a
menu
that
lets
you abort the test. Do not do
a
full
program break.
If
you
break
the
program, some
cal
constants may already have
been altered
in
the instrument.
A
controlled abort lets the program
replace the instrument's
original cal constants before exiting the test.
A
hard
STOP
does not do
this,
and when you run the program again, the program indicates:
DUT
CAL #AT10
OFFSET
=
### BUT
FILE
=
###
[OUT OK]
[NEITHER]
[FILE OK]
This screen
tells you
that
a
cal
constant
difference exists
between
the
data
disc image
file
and the
instrument.
It
then lets you choose which
is
the correct cal constant.
In
this
case,
the
image
file is
correct. For
each
cal
constant discrepancy
(sequentially) displayed, press
[FILE OK].
After several cal
constant discrepancies are displayed,
the program returns to normal operation.
NOTE:
The program repeats the Set IF attenuator statement for each measurement, even though
the
proper attenuation
is
set.
Procedure
1.
Select the desired softkey and respond to the program prompts:
[ADJUST] does
everything.
It
adjusts the calibration constants,
verifies
power accuracy
at
every
attenuator
step,
and sums the attenuator performance data
with the output
power
accuracy data,
comparing the
results to
instrument
specifications.
After you select this
key,
select
[ALL]
to adjust and measure
all
attenuator steps.
If
you select only
one step, no data
is
stored.
If
you
wish
to
calibrate the attenuator
first, and
then measure the
low output
power accuracy
performance, select [ADJUST].
[MEASURE]
is
a
subset of the
ADJUST
test.
MEASURE
verifies the low output power accuracy at
every attenuator step, and sums
the
attenuator performance data
with the output
power
accuracy
data, comparing the results to instrument specifications.
After you select
this key,
press
[ALL]
to
measure
all
the attenuator steps and store the data.
If
you
select only
one
step, no
data
is
stored.
Select the
MEASURE
softkey
if
you
only wish to verify the low output
power accuracy
perfor¬
mance.
[SPEC]
is
a
subset of the above tests.
SPEC
simply
takes
existing attenuator
performance data
and sums
it
with
output
power accuracy
data.
The
result
is
compared
to instrument
specifications
8-10
Tests/Utilities
Attenuator Calibration and Operation Verification
Test Software

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents