4-9.
SINGLE SIDEBAND PHASE NOISE (Cont'd)
7.
Adjust the HP
3585A marker
to the peak of the
IF.
8.
On
the
HP 3585A,
press
COUNTER.
9.
Adjust
the frequency of the LO
for an HP
3585A COUNTER
reading of
1
MHz ±0.9
Hz- This
removes
any
frequency offsets between
the
DUT
and LO.
10. Adjust the output
power
of
the DUT for
0
dBm at
the
peak of the
IF.
Select
a
REFERENCE
LEVEL
of
0
dBm
on the
HP 3585A.
SYSTEM CALIBRATION
30 Hz Offset
Calibration
(not applicable to units
equipped
with Option 007, Relaxed Phase Noise specifications)
11.
On the
HP 3585A,
press
FREQUENCY SPAN, 250
Hz, RES BW,
3
Hz. Allow one
sweep
to occur.
Ensure power
at
peak
if
IF
is still
0
dBm.
12.
Reduce
LO frequency by
30
Hz (IF
=
1.00003
MHz
±0.9
Hz).
13. Allow
one sweep of
the
HP 3585A
to occur. Adjust the HP
3585A marker
to the peak of the
IF.
Record the marker power
level (Cal Data) into Table 4-19.
Table 4-19.
Phase Noise
Calibration
Data
Step
13
14
16
19
21
22
Offset (Hz)
30
100
1K
10K
100K
0
Cal
Data
(dB)
CD=
CD
=
CD=
CD=
CD
=
IFDC=
100
Hz
Offset Calibration
14.
Reduce
the LO frequency by
70 Hz (IF
=
1.0001
MHz
±0.9
Hz) Repeat step
13.
1
kHz
Offset Calibration
15. On the HP
3585A, press FREQUENCY SPAN, 2500 Hz,
RES BW, 30 Hz.
16. Reduce LO
frequency
by
900 Hz
(IF
=
1.001
MHz
±0.9
Hz).
Repeat step
13.
10
kHz
Offset Calibration
17.
Connect
1
MHz
Notch Filter as shown
in
Figure 4-9. Select
1
MOhm
input impedance on the HP
3585A.
18.
On the HP 3585A,
press
FREQUENCY SPAN,
25 kHz, RES BW, 100 Hz.
19.
Reduce LO
frequency by
9
kHz
(IF
=
1.01
MHz
±0.9
Hz).
Repeat step