HP 8340B Manual page 55

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

4-10.
POWER SWEEP
TEST
Specification
Table 4-23.
Power Sweep Test Specifications
Power Sweep
Range
Displayed:
0
to
40 dB/sweep
Actual:
At least
10
dB
at any given
frequency (at least 20 dB
in
DECOUPLED
mode);
see Figure
2 in
the HP
8340B/41B
specification table.
Accuracy
Starting Power Level:
Same
as Output Power Accuracy
Power Sweep
Width and Linearity: See Figure
2
in
the HP
8340B/41B specification table.
In
normal operation
(a),
the
ALC does
not operate below —9.95
dBm
(see Figure
3 in
the
HP
8340B/41B
specification table), and so the maximum power
sweep
range
is
the difference of —9.95
dBm and the maximum leveled power available
at
the frequency range of Interest (specified leveled
power shown
in
diagram).
In
the DECOUPLED mode
(b), the
power sweep range
is
extended because
the ALC can operate down
to
—20 dBm.
Description
The DUT
is
set to allow the ALC and the step attenuator
to be
operated independently
([SHIFT] [PWR
SWP]). The DUT
is
then set to do
a
20 dB power
sweep
from —20 dBm to
0
dBm at
a
CW frequency.
The DUT
is
set
to do
a
manual sweep while
its
output power
is
measured
at the
two
end points, —20
dBm and
0
dBm,
using
a
power meter.
HP
8340B/41B
UNDER
TEST
POWER
METER
ADAPTER
APC-3.5(f)
to
N(m)
(6341B
ONLY)
POWER
SENSOR
OUTPUT
TEST
A
CAL
1
---—^•..——^
Figure 4-10.
Power
Sweep Test Setup
HP 8340B/41B
Performance Tests