5-4.
M/N
LOOP,
A32,
ASS, AND A32A1
(Cont'd)
6.
Connect DVM
to A32TP1
TUNE
test point. Refer
to Figure
5-8
for location of A32 assembly.
7.
Loosen locknut
on A32A2C1
FREQ ADJ control and
adjust
A32A2C1
for
a
DVM
reading of
-35.0
Vdc
±0.5
Vdc. Retighten A32A2C1 locknut. Verify that
DVM
reading
is still
within
tolerance.
8. Press
[CW]
on the
DUT,
then enter
[2]
[3] [0] [0]
[MHz].
9.
Turn the Spectrum Analyzer's
Signal
Track
off and then set the Spectrum Analyzer for
a
Center
Frequency
of 179.230
MHz.
Turn the Signal
Track
on. The
frequency
indicated on Spectrum
Analyzer should
be
179.230
MHz ±1
count and the
DVM
should read
-2.8
Vdc
±
1.0
Vdc.
10.
Set the DUT's LINE
switch
to STANDBY.
11. Disconnect the
DVM from the A32
assembly. Remove
the
A32/A33
assembly
from the instrument
and place
it
on
an
extender board.
12. Disconnect the
A32 output
cable A32A1
W1
from A33J3 (Figure 5-8) and
connect
this cable using
SMB
male-to-male adapter to the input
of the
Spectrum Analyzer. Set the DUT's LINE switch
to
ON
and
press
[INSTR
PRESET].
13.
Turn the Spectrum Analyzer's Signal Track
off and then set the Spectrum
Analyzer Center
Fre¬
quency to
375 MHz,
the Frequency Span
to 100
MHz, Log Scale
5dB/Division, and the
Reference
Level to
+5
dBm.
I
CAUTION
)
ViM^MfM^M
Do not apply
a
positive voltage to A32TP1 or damage may occur to
the
VCO
tuning diodes.
14. Connect
the low-voltage power
supply to the
OUT as follows: Positive lead to ground (do
this
first).
Negative lead to A32TP1 TUNE test
point.
Set
the
output
of the supply for
-35.0
Vdc
±0.5
Vdc.
15.
Adjust A32A2C5 PWR for
a
VCO
output level of
0
dBm
±
2
dB as indicated on Spectrum
Analyzer. Refer
to Figure
5-8
for location of adjustment.
16. Slowly reduce the voltage output of the external low-voltage power supply connected
to
A32TP1
TUNE
test point while monitoring the VCO output
level
on the
Spectrum Analyzer
and voltage
level on
DVM.
17.
The VCO
output
level should be
greater than
-2
dBm between 395 MHz
(-35
Vdc) and
355 MHz
(-2.8
Vdc).
18.
Repeat Steps
2
through
9
to check frequency accuracy.
19.
Set
the
DUT's LINE
switch to
STANDBY.
Disconnect
all
the test equipment from the
A32/
A33 assembly. Reconnect the cable
A32A1
W1
to A33J3.
20. Reinstall the A32/A33 M/N Output Assembly and remove the jumper from A59TP4
to
A59TP5 (UNLK
indicator disable).
21.
Disconnect the cable from the DUT's rear-panel EXT
FREQUENCY STANDARD,
recon¬
nect the BNC cable between the INT and EXT connectors, and set the adjacent switch
to
INT.
HP8340B/41B
Adjustments