5-15.
ALC ADJUSTMENTS (Cont'd)
ADJUST
AM OFFSET
27. Connect
the
Power Sensor
to the
DUT's RF OUTPUT.
Press
[INSTR PRESET]. Press [CW] and
enter
[1] [GHz], The
power
level should be 0
dBm. Disconnect any AM
inputs.
Access Calibration
Constant Number 43 by entering the following: [SHIFT]
[GHz] [4] [3] [Hz] [SHIFT] [MHz] [1] [2]
[Hz] [SHIFT] [kHz] [2] [2] [Hz].
Adjust Calibration Constant
Number 43
for no
more than 0.05
dB
change
in
power meter
indication
when turning AM on then
off.
AM
GAIN
ADJUSTMENT
28. Set up the following equipment as shown
in
Figure
5-50 with the power
supply
set
to
0
Vdc. Allow
the equipment to warm
up
for
at least
30
minutes.
OUT
POWER METER
0
b-
@
I
a
V
a
,ADAPTER
'APC-S.Sd)
to N(m)
BNC
TEE
Figure 5-50.
AM GAIN
Adjustment
Test
Setup
29. Calibrate the
power meter
and then set the calibration factor switch for
a
1.5
GHz measurement.
Connect
the
power sensor
to the
DUT's RF
output.
30. Press the following key
sequence
on the DUT:
[INSTR PRESET]
[CW]
[1] [.]
[5] [GHz]
[POWER LEVEL]
[-]
[5] [dBm]
[AM].
31. Adjust the power
supply for
0
Vdc. Press dB
[REF]
on the
power meter
(the
power
meter
will
now
indicate
the power
level
changes from
this
reference).
Adjustments
HP 8340B/41B