FM ACCURACY AND FLATNESS (Cont'd)
9. Press
[SHIFT]
[CW]
on the
DUT.
Note the
CW
frequency shown
in
the ENTRY DISPLAY. This
activates the
fine
tuning feature on the DUT.
The CW
frequency
will be
displayed
in
the
DUT's
ENTRY
DISPLAY.
Take
note of the flashing cursor
in
the
ENTRY
DISPLAY. Press the STEP UP or
STEP DOWN
keys to place the flashing cursor just to the left of the decimal
point.
Turn
the
DUT's
rotary knob (changing the CW frequency)
until
the
marker
on the HP
3585A's
CRT
dips. (It
is
normal for
a
spurious
signal to be
displayed below the
actual
marker. Keep
turning the
DUT's
knob
until
the
marker
dips.)
Press the STEP UP key twice. Turn the DUT's rotary knob
until
the marker on the HP
3585A
reaches minimum amplitude. This should
be possible every
75
MHz
or so due to the length of the
delay
line.
NOTE
Step
15
will instruct the operator to
set the DUT's FM
sensitivity to 10
MHz/V when
performing step
10. If
the
HP 355D step
attenuator
has not
been set
to 50 dB, the DUT's OVERMOD light
will
come
on.
10.
Move the output of the HP 8447F to the DUT's FM
input.
Turn FM ON and
AM
OFF. Verify that the
DUT's FM
sensitivity
is
set to
1
MHz/V
(default
condition).
11.
Set the HP
3585A
to
2
dB/DIV
and turn
A-B back ON. Press [CLEAR A]. Press [REF LVL] and
turn
the HP
3585A's
rotary knob (usually counter-clockwise)
until
the
marker
is
on the
center
horizon¬
tal graticule. If
the OVER
RANGE beep and message
occur, turn the rotary
knob
in
the opposite
direction. Don't
be
concerned
if
the marker does not move at
first,
it
may take several complete
turns of the rotary knob before
any
change
is
noticed on the
CRT.
12.
Press
[CONT] sweep
on the HP
3585A. The spectrum analyzer
will now
display the FM flatness
from 50 kHz to
10
MHz.
The
HP
3585A's marker offset feature
can be used to display
a
digital
readout of the
signal
power
level.
The readout
will be
expressed relative
to the
power
level at
100 kHz.
To
activate
this
feature,
press the following keys on the HP 3585A:
[MANUAL]
[1] [0]
[0] [kHz] [OFFSET]
[ENTER OFFSET] [MARKER]
The power
level at
any chosen frequency between 50 kHz
and 10
MHz
(relative
to
the 100
kHz
power
level)
will
be displayed
in
the
upper
right hand
corner
of the
CRT. Turn
the HP
3585A's
rotary knob
to
change the measurement frequency.
13.
Determine the maximum and minimum deviation
points from the 100
kHz
reference.
The
dif¬
ference between these
points should be :s3 dB. Record the
worst case power
level on the
Test
Record Card.
14.
Repeat Steps
6
through
13
for frequencies of 2.5,
6.9,13.4,
and 18 GHz.
If
testing an HP 8341B,
proceed to step
15.
If
testing an
HP 8340B,
also
perform the following: Press
[CW] [2] [0] [GHz]. Repeat
steps
6
through
13.
15.
Repeat Steps
6
through 14 with the HP 355D step attenuator set to 50 dB. When repeating step
10,
set the FM
sensitivity to
10
MHz/V,
not
1
MHz/V.
HP 8340B/41B
Performance