5-16.
LEVELED RF OUTPUT ADJUSTMENTS (Cont'd)
49. Use the step keys to step through Band
4.
If
squegging occurs, adjust A24R11 (X4C) slightly
CCW
to eliminate the squegging.
NOTE
Test
for squegging at power levels from maximum leveled power to —20
dBm
in 5
dB increments.
50. Press the DUT's
[SHIFT] [PWR SWP]
and enter the maximum ALC power
level
that
will
be
a
5
dB
increment below max leveled power
(i.e.,
5,0,
—5).
Repeat step 49
at this
power
level.
Enter the
next
5
dB increment and repeat step 49
until
the test for squegging has been performed from
maximum leveled power to —20 dBm.