5-10.
YO DELAY COMPENSATION, A54 (Cont'd)
BOTTOM VIEW
1
1
1M
——C
^
rU
X)
-^
L
^'
^
\
s
\
f
A54YO
PRETUNE
DAC/DLY
COMP
PW
A54R36
COFF
A54R32
CGN
A54R30
Figure
5-30.
YO
Delay
Compensation Adjustments Location
YO DELAY
ADJUSTMENT
4.
Connect the test equipment as shown
in
Figure 5-29.
5.
Set
the
DUT
as follows:
a.
Press [INSTR PRESET]
b.
Press [START FREQ] then enter
[6]
[.]
[9] [GHz]
c.
Press
[STOP FREQ]
then enter
[1] [3]
[.] [5]
[GHz]
d.
Press
[M1]
then
enter [7] [.] [2] [GHz]
e.
Press
[AMTD MKR]
on.
f.
Press
[SAVE]
then enter
[2]
g.
Press [SWEEP
TIME]
then enter
[2] [0] [0] [msec]
h.
Press
[SAVE]
then enter
[1]
i.
Press Frequency Marker [OFF]
j.
Press
[M2]
then enter
[1] [3] [.] [2] [GHz]
k.
Press
[SAVE]
then enter
[3]
I.
At ENTRY pad, press
[AUTO]
(sweep time)
m. Press
[SAVE]
then
enter [4]
n.
Press
[RECALL]
then
enter [1]