Chapter 8.
Calibration Constants Utility
8.0
Description
This
program
is
not
a
test
in
that
it
does not verify
conformance
to
a
specification- Rather
it
Is
a
utility
program which allows the operator to manipulate
the Calibration Constants
In
the
instrument or
In
the data disc
file.
Calibration Constants either from the Instrument or
data
disc
may be viewed or printed, Cal Constants may
be
transferred
from the instrument onto the
data disc (Image
file),
or vice versa; or the
Cal Constants
in
an
image
file
may be compared
with
those
in
the
instrument
8.1
Test Equipment
None required.
8.2
Procedure
1
The softkeys provided for
this
utility are as
follows
COMPARE
-
Compares
current Instrument Cal Constants with the image
file.
This
is
the key
you should
select
to
meet the requirements for
the opera¬
tion
verification
ReadDUT
-
Reads the DUT's Cal Constants
into
the computer's memory
GetFILE
-
Loads
the Cal Constant
Image
file
for the
OUT
into the
computer's
memory-
EXIT
-
Return
to
the test menu
Select
the desired softkey
2.
After any
of the
above three selections
are made, the program
will
offer
the
following
additional softkeys:
PRINT
-
Prints the Cal Constants currently
in
the computer's
memory
DISPLAY
-
Displays the Cal Constants currently
in
memory,
Wrt OUT
-
Writes
the Cal
Constants
in
the
computer's memory
to
the instru¬
ment's
WORKING
Cal Constant memory and then stores the Cal
Constants
into the
instrument's PROTECTED Cal Constant memory
SavFILE
-
Takes
the Cal
Constants
in
the computer's memory and creates
a
permanent image
file on the
data disc-
Select the desired softkey and respond to any prompts provided
by
the program-
HP
8340A/B
and
HP 8341A/B
Calibration Constants Utility
25/26