10.2 Procedure
1
The softkeys provided for
this
test are as follows:
Setup
-
Provides
a
graphical display of the
CW
Frequency Accuracy Test
setup
on the
CRT
M/N
-
Test
the
M/N
phase lock loop
only.
20/30
-
Test
the
20/30
phase lock loops only
ENDPNT ACCY
-
Test
the synthesizer's frequency endpolnts and the YIG oscillator's
frequency
endpolnts
ALL
-
Test
all
of the above- This
is
the key you should select to
meet the
requirements
of the
operation verification
Print
-
Must
be
pressed before
each test
is
run
to obtain
a
printout of the test
data No data
Is
stored for this test.
DONE
-
Return
to the
test
menu
Select
the
desired softkey and respond to
any prompts provided by the
program
In
order to
get
the
test menu
STATUS
to Indicate PASS,
all
three
of the
above test must
be
run
and pass
2.
If
a
test should
fail,
the program
will
pause and give you
a
softkey choice
of:
CONT.
-
Proceed
with the next
step of the test
REPEAT
-
Repeat
the test that was
run.
ABORT
-
Return
to
the softkey
menu
in
step
1
Select the desired softkey and respond
to any prompts provided by
the program-
so
CW
Frequency
Accuracy
Test
HP 8340A/B
and
HP 8341A/B

Advertisement