HP 8340B Manual page 237

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

5-5.
20/30
LOOP PHASE LOCK, A36, ASS, ASS, A40, AND
A43 (Cont'd)
49. Connect the
Function
Generator to A36TP3. Adjust the
Function
Generator amplitude
to
place the
20 kHz
signal displayed on the
Spectrum Analyzer
at
the
0
dBm reference
line.
50. Set the Function Generator to 50 kHz.
51.
Set
the Spectrum Analyzer
Marker at 50
kHz.
52. Adjust A36L7 and
A36L8
(50
kHz
NULL) to
null
the
50 kHz
signal displayed on
the Spectrum
Analyzer. Refer
to
Figure 5-12 for location of adjustments.
53. Note the level of the 50 kHz
signal. This level should be at least
65 dB down
from the level of the 20
kHz
response. Disconnect the test equipment from the
ASS
assembly.
54. Disconnect the active probe from the Spectrum
Analyzer
and connect the
Spectrum Analyzer
input to
A36J2
(OUT 200-300 MHz)
using
a
BNC
to
SMB
snap-on
test cable. Set the LINE switch
to
ON.
55.
Set
the
Spectrum Analyzer as
follows:
INSTR PRESET
................................
..............
Press
START FREQ
...................................................
150
MHz
STOP FREQ
...................................................
470 MHz
REF
LINE
....................................................
+10
dBm
MARKER
.................................................
PEAK SEARCH
56. Connect an external low-voltage
dc
power supply positive lead to A36TP3
and the negative lead
to
any convenient chassis ground.
Set
the
external power supply for
+16.0
Vdc
±
0.1
Vdc.
57.
The
Oscillator frequency should be 310
MHz
±10
MHz
as indicated
on the
Spectrum Analyzer
If
not, remove the metal shield from
A36
and
increase or
decrease
the spacing between turns of
A36L4
and increase or
decrease
the area of the single turn of
A36L5
to properly tune the
oscillator.
58.
Change
the
power supply voltage
to
+4.0
Vdc ±0.1 Vdc.
59.
The
Oscillator
frequency
should
drop
below
200
MHz
with
an
amplitude greater than —7 dBm.
60. Repeat Steps 56 through 59
if
necessary to meet requirements.
61.
Set
the LINE switch to STANDBY. Reinstall the metal
shield on
A36
if it
was
removed,
reinstall
A36
(PLL1
VCO)
in
the instrument, and reconnect
all
cables.
62. Remove A38 (PLL1 IF) from the instrument and
install
it
on an
extender
board. Connect
all
the
cables to A38.
63. Set
the
LINE switch
to
ON
and press [INSTR
PRESET].
Press
[CW]
then enter
[1] [3]
[.]
[9] [7]
[MHz].
HP 8340B/41B
Adjustments