HP 8340B Manual page 242

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

5-7.
YO MAIN DRIVER,
ASS
Reference
Performance
Test: None
Service Section:
Sweep Generator
YO
Loop
Description
The Main
Coil Driver circuit gain and offset are adjusted for
the correct output frequency.
OUT
son
LOAD
wsi
U
A45J3
J
(OUTPUT
OF
A45
DIRECTIONAL
COUPLER)
FREQUENCY COUNTER
Figure 5-15.
YO Main
Driver
ASS
Test
Setup
Equipment
Frequency Counter
........
50Q
Termination, SMB(f)
...
Adapter, SMB (m) to SMB (m)
......
HP
5343A
HP
P/N 1250-0676
HP P/N 1250-0669
Procedure
1.
Position the
DUT
in
the test
position as shown
in
Figure 5-15 with top and bottom covers removed.
Remove W3
(Figure 5-17) and connect the frequency counter to
W51 as shown
in
Figure 5-15.
Allow one-half hour for
warm
up
time.
2.
Remove cable W38 from A49J2 and connect
a
50 Ohm load
to the end
of
the cable.
3.
Set the DUT as follows:
a.
Press
[INSTR PRESET]
b.
Press
[CW]
then
enter
[2] [.]
[3] [GHz].
4. Adjust
A55R47 (OFFSET,
Figure 5-16) for an indication of 2.3
GHz ±1 MHz
on the
Frequency
Counter.
5.
Press the
DUT's [CW]
key, then enter [6] [.]
[9] [9] [9] [GHz].
6.
Adjust A55R4 (GAIN, Figure 5-16) for
an indication
of 6.999 GHz
±1 MHz
on
the Frequency
Counter.
HP 8340B/41B
Adjustments

Advertisement

Table of Contents
loading

This manual is also suitable for:

8341b

Table of Contents