TEST DEPENDENCE
FLOW CHART
The
flow chart
in
Figure
4-1
shows test interdependence. As shown
in
the
flow
chart, you
must
perform the power sensor cal,
level
1
before you
run the
output power accuracy test,
level
2, and
that
you
must perform
the IF
attenuator
cal, level
1
before you run the low output
power accuracy
test,
In
addition,
you must perform the output power accuracy test before
you run
the tests
in
level 3.
The
tests listed
in
Table
4-1
but not included
in
the flow chart may be performed
in
any order.
LEVEL
1
Power Sensor
Cal
(
CQl
Utility
Softkey)
IF
Attenuator
Cal
( Cal
Utility
Softkey)
LEVEL
2
Output
Power Accuracy
LEVEL
3
Maxim
Leveled
um
Power
RF
Attenuator
Calibration
Figure 4-1.
Test
Dependence
Flow Chart
Test Dependency
Attenuator Calibration and Operation Verification Test Software