5-1.
+
22 VDC
POWER SUPPLY, A35
Reference
Performance Test: None
Service Section: Power Supplies
Description
The
+22V
supply
is
adjusted for
+22
Vdc.
DIGITAL
VOLTMETER
OUT
A35TP1
Figure 5-1.
+22
Volt
Power
Supply
Adjustments
Setup
Equipment
Digital
Voltmeter
DUM
...........................................
HP
3456A
Procedure
1.
Position
DUT
in
the test position and connect
equipment as shown
in
Figure 5-1.
2. Set LINE switch to
ON.
3.
The
yellow
+22V
indicator on A35 should be
lit.
4.
The
DVM
indication should be
+22.000 ±0.010
Vdc.
If
the indication
is
out of tolerance, adjust
A35R3
+22
ADJ
(Figure 5-2) control
for
the specified voltage.