Running the Cal Utilities
6.0
Introduction
The
Cal Utilities portion
of the software
is
provided
to
assist the operator
in
generating
a
calibration data
file
for the specific
HP
355D step attenuator and HP 8485A power sensor
being used during the tests- As mentioned
in
Chapter
4,
Test Dependency,
several of the tests
require the use of the data stored
in
these data
files
before they can be
run.
The data
files
generated by the operator are stored and accessed on the storage media whose address
is
defined by Calfactor_dir$ and
Cf_msus$
(discussed
in
Chapter
3).
When
using
unmodified
software and
either an HP
9122D
or
9133H, the data
is
located on drive
0
Before
running the Cal Utilities, you should obtain the
calibration data for the attenuator
and
power
sensor you
will
be using during the tests. Once the data
is
obtained,
you can
proceed
with the appropriate
utility.
Upon completion of
the
utility.,
the data you entered
will
be stored
The
only time you
will
need to re-run the
Cal Utilities
is
when the attenuator or power sensor
is
re-calibrated or
if
you
are
using
a
device for which
you
have
not yet
entered data
To
run the Cal Utilities, press the Cat Utility softkey.
You
will
then be given
a
choice of entering
the attenuator's calibration data (IF Attenuator softkey) or the
power sensor's
calibration data
(Pwr Sensor Cal softkey). The Exit softkey
will
return you to the
8340 FAMILY TEST SYSTEM
menu.
Make
your selection and proceed
to
the appropriate entry paragraph below.
6.1
Entry of
the
IF
Attenuator's
(HP
355D) Calibration Data
Press
the
IF Attenuator softkey to begin entry of
the
attenuator's
calibration data.
A
softkey
menu
which provides you the options of
Start
or Return
will
be displayed Press the Start
softkey
to begin entry of the calibration data or the Return
softkey
to return to the
CALIBRATION UTILITIES
menu
When
the Start softkey
Is
selected,
the program will
request
the serial
number of the
attenuator and then the actual value of attenuation for the
10
dB through 70 dB steps When
entering
the attenuator step values, enter
the actual
value (i.e,
if
the
10db
step
is
10.02, enter
10..02). Also, the
values entered should
be positive After entering
all
of the data,
you
will
be
given the
opportunity to make corrections
if
necessary. Answer the rest
of the
questions
supplied
In
the program and
then press Return to
return
to the
CALIBRATION UTILITIES
menu-
6.2
Entry of the
Power Sensor's
(HP
8485A) Calibration Data
Press the Pwr
Sensor
Cal softkey to
begin entry of the
power sensor's calibration
data.
A
softkey menu which provides you the options of Start or Return
will
be displayed, Press the
Start softkey
to begin entry of the calibration
data
or the
Return softkey
to return to
the
CALIBRATION UTILITIES menu Before
pressing the Start softkey, perform the 10
MHz
Cal¬
ibration
Procedure described
in
paragraph 6.3
HP
8340A/B
and HP 8341A/B
Running the Cal Utilities