Adjustments
INTRODUCTION
This section provides adjustment procedures for the
HP 8340B/41B
Synthesized
Sweepers. These
procedures should not be performed as routine maintenance but should be used (1) after replacement
of
a
part or component, or
(2)
when performance tests show that
the instrument's specifications
cannot be met. Before attempting any adjustment, allow
1
hour warm-up time for
the instrument.
Table
5-1 lists
the adjustment procedures by test number
in
the
order
they
appear
in
this
Section. Table 5-2
lists all
factory selected components.
Table
5-3
lists
adjustment procedures that interact between
assemblies. This table
lists
the
adjustment paragraphs
that must be checked when an
assembly
is
adjusted, parts replaced, or the assembly replaced.
SAFETY CONSIDERATIONS
Although
this
instrument has been designed
in
accordance with international safety standards,
this
manual contains information, cautions,
and warnings which must be followed to ensure safe opera¬
tion and to retain the instrument
in
safe
condition.
Service and adjustments should
be
performed only
by
qualified
service personnel.
WARNING
Adjustments
in
this section are performed with power supplied to the instrument while
protective covers are removed. There are voltages at many points
in
the instrument
which can,
if
contacted,
cause
personal injury or
death.
Be
extremely careful.
Adust-
merits should be
performed
only by trained service personnel.
Power
is still
applied to
this instrument with
the LINE switch set
to
STANDBY. There
is
no
OFF
position on
the LINE
switch.
Before removing
or installing any assembly
or printed
circuit board,
remove the power
cord from
the rear of
the instrument.
Capacitors inside the instrument may
still
be charged, even
if
the instrument has been
disconnected
from
its
source of ac power. Use
a
non-metallic
adjustment
tool
whenever
possible.
EQUIPMENT REQUIRED
Table
4-1 lists
the equipment required for
the adjustment
procedures.
If
the test equipment
recom¬
mended
is
not available,
other equipment may
be
used
if its
performance meets the "Critical
Specifi¬
cations" listed
in
the table. The test setup
used for an
adjustment procedure
is
referenced
in
each
procedure.
HP 8340B/41B
Adjustments
5-1