Troubleshooting
Logs displayed on the KION Monitor
Diagnostic log
To display the Diagnostic Log on the KION Monitor,
select:
Menu / Monitor Setup / Biomed / Logs / Diagnostic Log
An example of a Diagnostic Log, as shown on the
KION Monitor, can be found in the illustration below.
CPS/IDS Diagnostic Log
To display the CPS/IDS Diagnostic Log on the KION
Monitor, select:
Menu / Monitor Setup / Biomed / Logs / CPS/IDS Diag. Log
An example of a CPS/IDS Diagnostic Log, as shown
on the KION Monitor, can be found in the illustration
below.
Log content
The events are displayed in order Newest to Oldest.
The following information is shown in the logs:
• Total Number Of Diagnostic Events: Number of
diagnostic events in the memory.
• SW Version:
– Diagnostic Log: Software version installed in the
KION Monitor.
– CPS/IDS Diagnostic Log: Software version
6
installed in the CPS / IDS.
• Product Serial #:
– Diagnostic Log: KION Monitor serial number.
– CPS/IDS Diagnostic Log: CPS/IDS serial number.
• Diagnostic Number: Can be used to identify some
multigas analyzer logs.
• Class: Not intended for service purpose.
• Time: Time stamp (h:min) of the event.
• Date: Date stamp (month/day/year) of the event.
• Dup Cnt: Counts if a log is repeted several times
within a short time frame without interruption of
other logs. Only the first event is time stamped and
shown. The repeted logs will only enumerate this
Dup Cnt. Example: Dup Cnt 6 shows that this log
has been repeted 6 times after the first event, i.e.
there is a total number of 7 logs.
Note: Dup Cnt is used only in the Diagnostic Log.
In the CPS/IDS Diagnostic Log, also repeted logs
are shown and time stamped.
6 - 20
• Line #: Not intended for service purpose.
• Description:
– Diagnostic Log: Log description written in
decimal numbers.
– CPS/IDS Diagnostic Log: Log description written
in hexadecimal numbers containing:
– Log number. 4 characters to be found directly
after the hex-number 55.
– Sub data. All 10 characters to be found directly
after the Log number. In most of the logs, only
the 4 last subdata characters are used. Only
logs 120 and 185 uses all 10 characters.
Log storage capacity
SC 7000 / 9000XL: .................................... 25 events
SC 9000: ................................................... 100 events
CPS / IDS: ................................................. 100 events
Siemens-Elema AB