295
... 01 27 ...
CON:
DIAG_EMERGENCY_TOO_LONG_INSP_PRS
Sub data:
... AB CD
INSP_TRANSDUCER_PRESSURE
x
296
... 01 28 ...
CON:
DIAG_EMERGENCY_TOO_LONG_EXP_PRS
Sub data:
... AB CD
EXP_TRANSDUCER_PRESSURE
x
297
... 01 29 ...
CON: DIAG_EMERGENCY_TOO_LONG_ISYN
Sub data:
... AB CD
EXP_TRANSDUCER_PRESSURE
INSP_TRANSDUCER_PRESSURE
UPPER_PRESSURE_LIMIT_SETTING
Fault: The system is not evacuated within 2 seconds when a high pressure
emergency situation occurred.
Description: The measured INSP pressure when the emergency situation occurred
can be read in sub data.
What to do:
1.
Check that EVAC system is not blocked.
2.
Redo a PuC to check that MV2 and PEEP valve works properly.
3.
Perform a Calibration
4.
Report to EM-HSC.
Shows the measured pressure (1/10 cmH2O) for the INSP pressure transducer in
hexadecimal.
Fault: The system is not evacuated within 2 seconds when a high pressure
emergency situation occurred.
Description: The measured EXP pressure when the emergency situation occurred
can be read in sub data.
What to do:
1.
Check that EVAC system is not blocked.
2.
Redo a PuC to check that MV2 and PEEP valve works properly.
3.
Perform a Calibration
4.
Report to EM-HSC.
Shows the measured pressure (1/10 cmH2O) for the EXP pressure transducer in
hexadecimal.
Fault: The system is not evacuated within 2 seconds when a high pressure
emergency situation occurred.
Description: Three consecutive diagnostic logs. The first log shows the pressure
measured on the EXP pressure transducer, the second shows the INSP transducer
and the third shows the UPL knob setting. Refer to sub data.
What to do:
1.
Check that EVAC system is not blocked.
2.
Redo a PuC to check that MV2 and PEEP valve works properly.
3.
Perform a Calibration
4.
Report to EM-HSC.
Shows the measured pressure (1/10 cmH2O) for the EXP pressure transducer in
hexadecimal.
Shows the measured pressure (1/10 cmH2O) for the INSP pressure transducer in
hexadecimal.
Shows UPL knob setting (1/10 cmH2O) in hexadecimal.