CON: DIAG_ HERC_TIMEOUT
x
232
... 00 E8 ...
CON:
DIAG_PUC_BIBD_INTERNAL_LEAKAGE
Sub data:
... 00 02
DGBD_WAIT_FOR_EMPTY_BAG
... 00 05
DGBD_WAIT_FOR_FULL_BAG
... 00 0B
DGBD_WAIT_FOR_FULL_BAG_FINAL_
CHECK
x
235
... 00 EB ...
CON:
DIAG_TIMETEST_CURRENT_MAX_LOAD
x
236
... 00 EC ...
CON: DIAG_TIMETEST_TOTAL_MAX_LOAD
Fault: KION Monitor do not respond during PuC or Calibration.
Description: The KION Monitor do not respond to commands within one second when
performing a PuC or Calibration.
What to do:
1.
Change KION Monitor and check if error disappears
2.
Check the function of the monitor docking station e.g by connecting an
external docking station to P83 on PC 1753 in the Battery drawer.
3.
Check/replace the CPS/IDS board.
Fault: Leakage detected or one of the pressure transducers is not working properly
during the Bellows position sensor calibration in the PuC.
Description: Refer to sub data below
What to do: Refer to sub data below.
The Breathing bellows is emptied with a drive gas flow of 30 l/min.
The Breathing bellows is not detected to be in lowest position after 30 seconds, i.e. the
PEEP Pressure - Insp. pressure < 17.5 cmH2O.
What to do:
1.
Check the seal on the bellows base plate.
2.
Check/replace EMV 7.
3.
Check/replace the Drive gas module.
The Breathing bellows is filled with a flow of 27 l/min from the AIR/N2O or O2 gas
modules.
The Breathing bellows is not in top position after 30 seconds, i.e. the Exp. Pressure -
PEEP Pressure < 15 cmH2O.
What to do: Check/replace the Breathing bellows (leakage).
A final check that the calibrated Bellows position sensor is working by moving the
bellows to top position again.
The Breathing bellows is filled with a flow of 27 l/min from the AIR/N2O or O2 gas
modules.
The Breathing bellows is not detected to be in top position after 30 seconds, i.e. the
Exp. Pressure - PEEP Pressure < 15 cmH2O.
What to do: Check/replace the Breathing bellows (leakage).
Description : Incorrect SW version.
What to do: Check the CON-SW version in component log and report to EM-HSC.
Description : Incorrect SW version.
What to do: Check the CON-SW version in component log and report to EM-HSC.