Troubleshooting
Leakage exceeds 6 l/min – Technical error codes
The error is recorded as Diagnostic Log 261 alt. 263.
KION Monitor – A four digit code 990X, shown to the
left of the leakage check texts in the PuC window.
Recommended actions
Code 9901: Software loading failure:
Problem when loading / reading the software in
KION/KION-i.
1. Power-cycle KION/KION-i.
6
2. Replace PC 1730 Control.
Code 9902: Leakage >19 l/min (disconnect):
Test pressure was not obtained within 15 s; the
leakage check fails. Obtained pressure is shown in
the digital display "Fresh gas flow".
1. Make sure that the basic leakage checks are
performed. Refer to the Operating Manual,
chapter "Troubleshooting"
2. Check for leakages in the workstation. Refer to
section "Check for leakage" in this chapter.
Code 9903: Software loading failure:
Start pressure mean value cannot be calculated.
1. Power-cycle KION/KION-i.
2. Replace PC 1730 Control.
6 - 4
KION/KION-i control panel – A single digit code 1-9,
stated in the digital display for CMV. The digit corres-
ponds to the 4th digit in the technical error code.
Code 9904: Software loading failure:
End pressure mean value cannot be calculated.
1. Power-cycle KION/KION-i.
2. Replace PC 1730 Control.
Code 9905: Gas leakage into the workstation:
End pressure more than 1 cmH
pressure.
1. Make sure that the basic leakage checks are
performed. Refer to the Operating Manual,
chapter "Troubleshooting"
2. Check for leakages into the workstation. Refer to
section "Leakage into the system" in this
chapter.
Siemens-Elema AB
KION / KION-i
O above Start
2
E382 E392E 061 01 02 02