Siemens KION Service Manual page 163

Anesthesia workstation
Table of Contents

Advertisement

x
417
... 01 A1 ...
x
419
... 01 A3 ...
x
420
... 01 A4 ...
x
421
... 01 A5 ...
100
x
x
601
... 02 59 ...
560
x
x
602
... 02 5A ...
560
x
x
603
... 02 5B ...
560
MEA: DIAG_AD_ERROR_DOUBLE_FAULT
MEA: DIAG_INTERNAL_ERROR
MEA: DIAG_NO_DATA_FROM_MGM
MEA: DIAG_NO_INSP_SIGNAL_FROM_CON
MGI: DIAG_CPU_ERROR
MGI: DIAG_ROM_ERROR
MGI: DIAG_RAM_ERROR
Fault: Hardware failure.
Description: Failure to read A/D-port twice.
What to do: Replace PC 1737 Measuring.
Fault: Internal program failure.
Description: Diagnostic log regarding faulty states in MEA software.
What to do:
1.
Replace PC 1737 Measuring.
2.
Report to EM-HSC.
Fault: The CAN/RS232 interface reports an error.
Description: The multigas analyzer is not reporting concentration data as presumed.
Note: This diagnostic log may appear now and then, but will impact functionality only if
repeted several times in a row.
What to do (Only if repeted):
1.
Check if the multigas analyzer is displaying concentration values correctly
on the KION Monitor.
2.
Troubleshoot the multigas analyzer:
– Function check.
– Calibration.
– Multigas analyzer service software.
Fault: No INSP or EXP signal is received from CON.
Description: The signal informing that a new expiration or inspiration has started has
not been received from the CON within 30 seconds.
This diagnostic log can only come in controlled ventilation modes.
What to do:
1.
Replace PC 1737 Measuring.
2.
Replace PC 1730 Control.
3.
Report to EM-HSC.
Fault: Hardware failure.
Description: CPU is not working properly during startup hardware test. Alarm 560 is
generated.
What to do: Replace PC 1747 CAN/RS232 Interface.
Fault: Hardware failure.
Description: ROM is not working properly during startup hardware test. Alarm 560 is
generated.
What to do: Replace PC 1747 CAN/RS232 Interface.
Fault: Hardware failure.
Description: RAM is not working properly during startup hardware test. Alarm 560 is
generated.
What to do: Replace PC 1747 CAN/RS232 Interface.

Advertisement