• When mounting the bellows base plate, make sure
that the bellows housing seal (5) on the bellows
base plate is correctly mounted, see illustration.
• When mounting the bellows position sensor (6),
make sure that the sensor is adjusted into correct
position (level) as shown in the illustration.
Support plate for patient cassette
• Remove the bellows base plate as described
above.
• Disconnect the mushroom valve control gas supply
tube (1). If necessary, cut the tube.
• Disconnect N77 (2) and N78 (3) from PC 1755
P
C
V
C
ATIENT
ASSETTE
ALVE
• From underneath the patient unit, remove the
screws (4) and lift off the absorber support plate (5)
for the CO
Absorber.
2
Note: Hold the absorber support plate when
removing the screws. The absorber support plate
will fall down when the screws are removed.
When mounting the absorber support plate:
• Apply special grease on the hinges.
• Secure the screws (4) with Loctite 243.
Disassembling and assembling
.
ONTROL
4 - 17