KION / KION-i
Diagnostic log groups
The diagnostic logs can be divided into four main
groups:
1. Technical Alarm logs.
2. Pre-use check logs.
3. Calibration logs.
4. Other diagnostic logs.
All logs are listed in the "Diagnostic Log table" below
in this chapter.
Technical Alarm logs
In the Diagnostic Log-example shown on page 21,
the two logs (27) with identical Time and Date stamp
indicates that "MEA is not responding".
The corresponding logs in the CPS/IDS Diagnostic
Log-example are ...001B....
If a Technical Alarm occurs, the first recommended
remedy is to power-cycle the KION / KION-i
workstation.
Together with the Technical alarm logs, there are
some "Additional logs" created, see below. These
Additional logs are shown in combination with the
Technical alarm logs. In the Diagnostic Log-example
shown on page 21, the two logs (27) are combined
with the Additional log "KION CRITICAL FAIL (100)
Source (0)" with identical Time and Date stamp
Pre-use check logs
Logs that indicates errors and events during the Pre-
use Check.
Calibration logs:
Logs that indicates errors and events during the
Calibration.
Other diagnostics:
Logs that indicates errors and events during startup
and operation of the KION / KION-i workstation.
E382 E392E 061 01 02 02
Additional logs
Together with the Technical alarm logs, there are
some additional logs created. These additional logs
are always shown in combination with the Technical
alarm logs.
The additional logs are only shown in the Diagnostic
Log, not in the CPS/IDS Diagnostic Log.
The additional logs are:
KION CRITICAL FAIL (100)
Technical Alarm followed by one or several clarifying
diagnostic messages.
KION (510)
Technical Voltage Alarm from ALA followed by one of
the clarifying diagnostic messages:
• 026 – ALA: DIAG_VOLTAGE_ERROR_24V_ON
• 042 – ALA: DIAG_VOLTAGE_ERROR_PLUS_15V_ON
• 043 – ALA: DIAG_VOLTAGE_ERROR_MINUS_15V_ON
MultiGas Failure (560)
Technical HW Alarm from the CAN/RS232 interface
followed by one of the clarifying diagnostic
messages:
• 601 – MGI: DIAG_CPU_ERROR
• 602 – MGI: DIAG_ROM_ERROR
• 603 – MGI: DIAG_RAM_ERROR
• 611 – MGI: DIAG_TIMER_ERROR
• 612 – MGI: DIAG_WD_ERROR