Service procedures
Software installation
There are a number of different softwares included in
the KION / KION-i workstation. Perform as follows
when updating these softwares:
CON-SW
The CON-software (CON-SW) is stored in a FLASH-
PROM on PC 1730 C
CON-SW can be installed either by replacing PC 1730
or by using the Software download tool. Instructions
for the software download are enclosed with the
download tool.
MEA-PROM
The MEA-software is stored in an exchangable pro-
gram memory (MEA-PROM) on PC 1737 M
To update the MEA-software, replace the MEA-
PROM.
ALA-PROM
5
The ALA-software is stored in an exchangable
program memory (ALA-PROM) on PC 1672 C
P
. To update the ALA-software, replace the ALA-
ANEL
PROM.
PAN-PROM
The PAN-software is stored in an exchangable
program memory (PAN-PROM) on PC 1672 C
P
. To update the PAN-software, replace the PAN-
ANEL
PROM.
MGI-PROM
The MGI-software is stored in an exchangable
program memory (MGI-PROM) on PC 1747 CAN/
RS232 I
. To update the MGI-software,
NTERFACE
replace the MGI-PROM.
KION MONITOR-SW
The KION MONITOR-software is stored in the KION
Monitor. New software can be installed from a PC
Card using the card reader on the KION Monitor.
Perform the software installation as described in the
instructions delivered with the PC Card.
5 - 16
. New versions of the
ONTROL
EASURING
ONTROL
ONTROL
CPS-SW alt. IDS-SW
The CPS-SW alt. IDS-SW is stored in the
C
I
OMMUNICATION
Network PCB). New software can be installed from a
PC Card using the card reader on the KION Monitor.
Perform the software installation as described in the
instructions delivered with the PC Card.
Configure the CPS / IDS
When delivered, the KION / KION-i workstation is
configured as a stand-alone unit. A configuration of
the CPS/IDS network functions is necessary to
perform e. g.:
• When the KION / KION-i workstation will be
connected to an I
• When installing a new CPS alt. IDS Network PCB
.
into a KION / KION-i workstation, stand-alone as
well as network connected. The default
configuration may not work correctly together with
the KION Monitor. If so, a "No Screen" alarm will
be activated on the KION control panel and the
messages "CPS/IDS Access failed" and
"OFFLINE" will be shown on the KION Monitor.
Furthermore, there will be no access to the CPS/
IDS functions in the KION Monitor service menu.
A complete description of a CPS/IDS configuration
for I
Network can be found in the "I
NFINITY
Network – Planning, design and installation
handbook".
The brief descriptions below shows the main steps
of a CPS/IDS configuration.
A PC with the software HyperTerminal (included in
Windows) can be used.
• Make sure that the PC and the KION / KION-i
workstation is turned Off.
• Connect an RS232 cable (Order No. 47 14 346
E530U to a COM port on the PC.
• Connect the other end of the cable to the X4 port
on the KION / KION-i workstation.
• Turn the PC On and the KION / KION-i workstation
to Standby.
• Start the terminal program and make the following
COM port-settings:
Bits per second: 19200
Data bits:
Parity:
Stop bits:
Flow control:
Siemens-Elema AB
(CPS Network PCB alt. IDS
NTERFACE
network.
NFINITY
8
None
1
None
E382 E392E 061 01 02 02
KION / KION-i