Siemens KION Service Manual page 17

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i
Monitor (MON)
The Monitor (MON) subsystem includes the:
• C
, CPS alt. IDS Network PCB.
OMMUNICATION INTERFACE
• PC 1754 C
alt. PC 1856 C
ONNECTOR
• PC 1761 MIB C
ONNECTOR
C
2.
ONNECTOR
• KION Monitor.
The C
OMMUNICATION INTERFACE
Siemens Communication Power Supply (CPS) alt.
Infinity Docking Station (IDS) module.
The software in the C
OMMUNICATION INTERFACE
SW alt. IDS-SW, and KION MONITOR-SW in the
KION Monitor is installed from a PC Card via the
KION Monitor.
The main functions of the MON subsystem are:
• Displaying patient parameters and wave forms.
• Displaying gas supply pressures.
• Displaying set and measured gas concentrations.
• Displaying alarms.
• Handling external communication.
Control (CON)
The CON-software (CON-SW) is stored in a FLASH-
PROM on PC 1730 C
ONTROL
CON-SW can be installed either by replacing PC 1730
or by using the Software download tool.
The main functions of the CON subsystem are:
• Control of breathing systems.
• Control of ventilation modes.
• Control and measuring of fresh gas flow including
gas mixture selection (O
• Calculation of inspiratory flow.
• Calculation of spontaneous breathing frequency.
• Conversion of panel settings and feedback to
panel.
• Controlling the "Pre-use check" and the
"Calibration" sequences.
• Control of all electromagnetic valves.
E382 E392E 061 01 02 02
.
ONNECTOR
alt. PC 1843 MIB
is also included in the
, CPS-
. New versions of the
/AIR, O
/N
O).
2
2
2
Siemens-Elema AB
Power supply (PWR)
The Power supply (PWR) subsystem includes the:
• AC/DC Converter.
• Internal battery.
• PC 1753 DC/DC C
ONVERTER
• PC 1675 P
S
.
OWER
UPPLY
• Isolation transformer (if connected).
• Control of power-up and power-down sequences.
The Power supply subsystem is responsible for
converting mains power and battery power to the
voltage levels requested by the system. Power
supply will also supply these voltages to the electrical
parts within the system.
Panel
PAN
Pneumatic
section
including
MEA
and
CON
+24 V_VALVE
Introduction
.
Multi gas
analyzer
interface
MGI
Monitor
MON
Alarm
ALA
+24 V
+12 V
-15 V
Power supply
-12 V
+15 V
PWR
+5 V
+13 V
GND
For supervision purposes
2 - 11
2