Siemens KION Service Manual page 139

Anesthesia workstation
Table of Contents

Advertisement

x
263
... 01 07 ...
CON:
DIAG_PUC_HANDVENT_LEAKAGE_FAILURE
Sub data:
... 26 AE
CAL_ERR_START_P_TOO_LOW
... 26 AF
CAL_ERR_NOT_ENOUGH_P1_SAMPLES
... 26 B0
CAL_ERR_NOT_ENOUGH_P2_SAMPLES
... 26 B1
CAL_ERR_P_HIGHER_AFTER_DELAY
... 26 B2
CAL_ERR_START_P_TOO_HIGH
... 26 B3
CAL_ERR_START_P_TOO_LOW_2
... 26 B5
CAL_ERR_P_NOT_RAISE_ON_REFILL
x
264
...01 08 ...
CON:
DIAG_PUC_HANDVENT_NOT_ACCEPTABLE
_LEAKAGE
Sub data:
... AB CD
LEAKAGE_MORE_THAN_150_ML
x
265
... 01 09 ...
CON:
DIAG_PUC_QCHECK_INSP_PRESS_NOT_RE
ACHED
Sub data:
... AB CD
QCHECK_FAILED_TO_BUILD_UP_PRESSUR
E
Fault: "Leakage manual ventilation" test failed during PuC.
Description: The test pressure is 25 cmH2O.
What to do: Refer to section "If the Pre-use check fails" in this chapter and to
information regarding "Leakage manual ventilation" in chapter "Service procedures",
section "Pre-use check".
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Refer to sub data for DIAG_PUC_AUTOVENT_LEAKAGE_FAILURE (261)
Fault: "Leakage manual ventilation" test failed during PuC. No error code stated.
Description: The leakage was above 150 ml/min
What to do: Refer to section "If the Pre-use check fails".
The leakage (ml/min) is presented in hexadecimal form.
Fault: "Checking pressure transducers" test failed during PuC.
Description: Failed to build up pressure in system to 10 cmH2O
(±4 cmH2O). MV1+MV5 are closed during this test sequence.
The pressure is slowly build up with a fresh gas flow of 6 l/min until the Insp. pressure
transducer measures 10 cmH2O. If the build-up pressure timeout of 10 s is reached,
this diagnostic log is written.
Refer to section "If the Pre-use check fails" in this chapter and to information regarding
"Checking pressure transducers" in chapter "Service procedures", section "Pre-use
check".
What to do:
1.
Perform a Calibration
2.
Check for leakage in the system, i.e. check that pressure transducer tubes
are properly connected.
3.
Replace PC 1750 Transducer (or the concerned PC 1781).
Reached build-up pressure (1/100 cmH2O) after 10 s measured by the insp. pressure
transducer in hexadecimal form.