... 00 0B
CAL_BIB_DETECTOR_CALIB_FAIL
Fault: Failed to calibrate Bellows position sensor.
Description: The difference between the E2-potentiometer value for empty bellows
and full bellows is less than 5 units. The difference value is shown in the Exp. minute
volume display.
What to do:
1.
Check that nothing is obstructing/disturbing the sensor.
2.
Check for dirt/cracks on the Bellows housing.
3.
Check that the Bellows position sensor is mounted in correct position (level).
Refer to section "If the Pre-use Check fails".
4.
Check/replace the Breathing bellows. The white ring on the bellows will not
reflect the light to the sensor properly.
5.
Replace PC 1766 Bellows position sensor.
6.
Replace PC 1750 Transducer.