Siemens KION Service Manual page 66

Anesthesia workstation
Table of Contents

Advertisement

Disassembling and assembling
4
6
H
4 - 20
<
;
9
6
Siemens-Elema AB
• Open the control panel housing (4).
• Disconnect the two panel cable connectors (5)
from P9 and P17 on the PC 1672 C
• Lift off the control panel housing (4).
When assembling the control panel housing, make
sure that there are no pinched cables. Defective
cables may short-circuit power supply, CAN-bus
signals, etc.
Front panel
• Remove the control panel housing as described
above.
• Remove the five screws (1) and lift off the front
cover (2).
• Disconnect and remove the alarm silence push
button (3).
• Remove all knobs from the potentiometers and
switches.
• The front panel is mounted to the control panel
housing with adhesive strips. Carefully remove the
front cover from the housing.
KION / KION-i
P
.
ONTROL
ANEL
E382 E392E 061 01 02 02