Siemens KION Service Manual page 20

Anesthesia workstation
Table of Contents

Advertisement

Description of functions
About this chapter
The Description of functions is divided in two levels:
• Basic principles – Basic principles containing a
survey of the function. This text is written with an
italic typeface. The text refers to the diagram
"Basic principles" in chapter "Diagrams".
• Enhanced information – Enhanced information
about the function. Normal typeface is used in this
text. This text refers to the following diagrams in
chapter "Diagrams":
– Pneumatic block diagram
– Electronic interconnection.
3
Text written in S
MALL
the diagrams mentioned above.
Battery drawer
The battery drawer contains functions for power
supply and communication.
1. AC/DC Converter
The AC/DC C
ONVERTER
rectifier and the transformer, is a complete unit that
is mounted on the Power & Communication interface
casing. The AC/DC C
to PC 1675 P
S
OWER
There are two mains power inlet fuses in the AC/DC
C
. The fuses and the AC/DC C
ONVERTER
are the only spare parts available for the AC/DC
C
.
ONVERTER
The AC/DC C
ONVERTER
converter compartment. If the fan fails, the signal
FAN_FAIL generates a diagnostic log message.
If the AC/DC C
ONVERTER
194°F), e. g. due to fan failure, the AC/DC C
will be disabled. The KION / KION-i workstation will
then be power supplied by the I
When the AC/DC C
ONVERTER
automatically be enabled.
Technical data for the AC/DC C
• AC/DC C
ONVERTER
• AC/DC C
ONVERTER
voltage level 21–30 V DC controlled by an analog
signal from the Charging/Discharging control block
on PC 1675 P
OWER
• Fuses: T 6.3 AL / 250 V AC (2 pcs).
3 - 2
C
refers to block names in
APS
, containing the mains power
is electrically connected
ONVERTER
.
UPPLY
ONVERTER
F
pulls cooling air into the
AN
is overheated (above +90°C /
ONVERTER
B
NTERNAL
ATTERY
cools down, it will
:
ONVERTER
input: 100 – 240 V AC, 50 – 60 Hz.
output: Max. 400 W, output
S
.
UPPLY
2. Internal battery
The I
B
NTERNAL
ATTERY
sealed lead acid batteries) supplies the KION / KION-i
workstation with backup power for min. 30 minutes
with full charged battery.
The I
B
NTERNAL
ATTERY
backup. The batteries are rated 12 V, approx. 24 Ah.
The I
B
NTERNAL
ATTERY
KION / KION-i workstation is connected to mains
power. Charging time for empty batteries is approx.
6 hours.
The actual I
NTERNAL
displayed on the KION Monitor. The battery capacity
indication, bargraph with %, is related to actual
battery voltage. "100%" is indicated when trickle
charge level is reached (fully charged).
The following applies concerning battery voltage
supply:
• Battery voltage supply is used as soon as the
AC/DC C
ONVERTER
voltage (e.g. if the power consumption is too high).
• If the AC/DC C
19 V (e.g. at loss of mains power), a "Switching to
battery" alarm indicates that battery voltage supply
is in use.
• First battery alarm "Limited battery capacity" is
activated when battery voltage reaches below 23 V
(22 V if an E
XTRA
These voltage levels correspond to "20%"
F
AN
indicated.
• Second battery alarm "No battery capacity" is
activated 10 minutes after first battery alarm or
when battery voltage is below 22 V (21.5 V if an
E
I
XTRA
NTERNAL
levels correspond to "0 %" indicated.
• All functions will be cut off when battery voltage
reaches below 21 V.
Battery capacity indication and alarms are generated
.
via the ALA power supervision function.
Service interval for the I
• The I
B
NTERNAL
"Twelve months maintenance".
• The I
B
NTERNAL
years. After replacement, allow the batteries to
recharge before clinical use of the KION / KION-i
workstation.
The lifetime of the I
used frequently to supply the KION / KION-i
workstation with operating power or if the battery
temperature is above +50°C (122°F) e. g. due to
B
D
ATTERY
RAWER
recommended to check the I
shorter intervals than 12 months.
Siemens-Elema AB
(two serial connected 12 V
is used as an internal 24 V
is continously charged when the
B
capacity is continously
ATTERY
voltage is less than the battery
output voltage is below
ONVERTER
I
B
is connected).
NTERNAL
ATTERY
B
is connected). These voltage
ATTERY
B
:
NTERNAL
ATTERY
must be checked during the
ATTERY
must be replaced every three
ATTERY
B
is reduced if it is
NTERNAL
ATTERY
F
failure. In such case, it is
AN
B
NTERNAL
ATTERY
E382 E392E 061 01 02 02
KION / KION-i