Service procedures
Connection with saved settings
• Start HyperTerminal using the HyperTerminal-
setting saved for communication with the KION /
KION-i. The HyperTerminal window will now be
open and active.
• Press the Enter key. The "System Console Main
Menu" will now appear in the HyperTerminal
window. If a sub-menu is displayed, return to the
"System Console Main Menu". Communication
with the KION / KION-i is established.
Disconnecting the PC from the KION / KION-i
• Make sure that the KION / KION-i is turned Off.
• Make sure that the PC is turned Off.
• Disconnect the RS232 cable between the PC and
the KION / KION-i.
5
5 - 20
Access the diagnostic logs from a PC
Downloading the diagnostic logs
The HyperTerminal program is used only as an
interface for communication with the KION / KION-i:
• Establish communication between the KION /
KION-i and the PC (see above).
• Enter selection 1 (Error Services) in the "System
Console Main Menu". Press the Enter key. The
"ERROR Driver Main Menu" will now appear.
• Select Transfer/Capture text in the HyperTerminal
menu.
• Select a name and a path for the target file.
• Click on Start. The target file is now ready to be
used.
• Enter selection 1 (View the Error Log) in the
"ERROR Driver Main Menu". Press the Enter key.
The log will now be copied to the target file.
• Wait until the copying is completed and the
"ERROR Driver Main Menu" appears again.
• Select Transfer/Capture text/Stop in the
HyperTerminal menu to close the target file.
• Exit the HyperTerminal program.
It is now possible to open the target file with a
suitable program, e.g. Notepad. To make it easier to
read the information in the log, it is recommended to
change the font (use e. g. Arial) and to delete
unnecessary parts of the file.
Deleting the diagnostic logs
• Establish communication between the KION /
KION-i and the PC (see description above).
• Enter selection 1 (Error Services) in the "System
Console Main Menu". Press the Enter key. The
"ERROR Driver Main Menu" will now appear.
• Enter selection 5 (Clear the Error Log) in the
"ERROR Driver Main Menu". Press the Enter key.
The log will now be deleted.
• Exit the HyperTerminal program.
Note: Do not delete the diagnostic logs, e.g. after
performed service. In case of future troubleshooting
on the workstation, as much relevant log information
as possible should be kept. The oldest logs will
automatically be overwritten when the log list is full.