Disassembling and assembling
9
4
;
4 - 16
H
6
<
INSP
EXP
MGA
MAN
Siemens-Elema AB

Patient unit

• Unclip and lift off the bellows housing (1).
• Lift off the breathing bellows (2).
• Lift off the bellows support plate (3).
• Remove the bellows base plate (4) including its
bellows housing seal (5) and the two connected
hoses.
Note: The two connected hosed (for drive gas and
for exhaust gas) must be disconnected inside the
pneumatic section.
The parts inside the patient unit are now accessible.
• This illustration shows the position of the tubes and
cables inside the patient unit. As it is possible to
rotate the breathing system, it is thus very
important to check tubes and cables when
assembling the patient unit. Make sure that no
tubes nor cables are squeezed when rotating the
breathing system.
• The detail drawing shows the position of the tube
nipples for the pressure transducer and multigas
analyzer tubes.