Siemens KION Service Manual page 89

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i
• Press Enter to access the "System console
menu". Make sure that the main menu "CPS
68030 SYSTEM CONSOLE" is displayed.
• Select "11" – Display/Modify NW Config Data
• Select "3" – Input all NW Config data
• Press "Enter" on all suggested settings (to accept
default settings) with the following exeptions:
– Bed label: Type e.g."BED1"
– CPS/IDS label: Type e.g. "OP1"
– Care Unit label: Type e.g. "OP"
– Monitoring unit label: Type e.g. "MON1"
– Hospital label: Type e.g. "Central".
It is not necessary but recommended to enter
"Recorder label". This makes it possible to connect a
recorder at any time. Type e.g. "REC1"as name of
the primary recorder as well as the secondary
recorder.
• Turn the PC and the KION / KION-i workstation Off.
• Disconnect the RS232 cable between the PC and
the KION / KION-i workstation.
• Restart the KION / KION-i workstation. The new
CPS configuration will now be used.
Display the CPS/IDS configuration
With a PC connected (as previously described) it is
also possible to display the CPS/IDS configuration.
• Access the "System console menu".
Make sure that the "CPS 68030 SYSTEM
CONSOLE" is displayed.
• Select "11" – Display/Modify NW Config Data
• Select "0" – Display NW Config
E382 E392E 061 01 02 02
The example below shows Network Configuration
data for a workstation configured as a stand-alone
unit:
CPS/IDS Host ID: 0
Monitoring unit ID: 1
IP Addr: 191.1.1.0
Subnet Mask: 255.255.0.0
Default Route: 0.0.0.0
Bed Label: BED1
Recorder 1 Label: REC1
Recorder 2 Label:
CPS Label: OP1
Monitoring Unit Label: M1
Care Unit Label: OP
Hospital Label: HOSP
Primary Recorder: REC1
Secondary Recorder: REC1
Primary Alarm Recorder: REC1
Secondary Alarm Recorder: REC1
Remote Silence Enabled: Yes
Remote Control Enabled: Yes
Alarm Group: 0
Will this bedside be monitored by a Central Station?: Yes
Network Mode: Standalone
Recorder Use: Global
Laser Printer IP Addr: 191.1.1.240
The example below shows Network Configuration
data for a workstation configured as a network node:
CPS Host ID: 55
Monitoring unit ID: 1
IP Addr: 191.1.1.55
Subnet Mask: 255.255.0.0
Default Route: 0.0.0.0
Bed Label: BED1
CPS/IDS Label: OP1
Monitoring Unit Label: M1
Care Unit Label: OP
Hospital Label: Central
Recorder 1 Label: REC1
Recorder 2 Label: REC1
Primary Recorder: REC1
Secondary Recorder: REC1
Primary Alarm Recorder: REC1
Secondary Alarm Recorder: REC1
Remote Silence Enabled: Yes
Remote Control Enabled: Yes
Alarm Group: 5
Will this bedside be monitored by a Central Station?: No
Network Mode: Networked
Recorder Use: Global
Laser Printer IP Addr: 191.1.1.240
Network Comm Stack Configuration: CPS/IDS Stack
NWeb Server Addr: 0.0.0.0
CPC/IDS Web Server Port: 80