ALA:
DIAG_MEASURE_NOT_FULLY_OPERATIONA
L
ALA:
DIAG_CONTROL_NOT_FULLY_OPERATIONA
L
ALA:
DIAG_PANEL_NOT_FULLY_OPERATIONAL
Fault: MEA is not in the same state as ALA.
Description: MEA is in Standby when ALA is in On (or vice versa) for more than 3.3
seconds. Technical alarm is generated.
What to do:
1.
Check if the Power switch is faulty and stuck between its fixed positions. If
the switch is sluggish, adjust the knob. If all diagnostics 038, 039, 040 are
shown in the log, this is a possible fault.
2.
Check that the MEA-PROM is correctly mounted
3.
Replace PC 1737 Measuring
4.
Replace PC 1672 Control Panel. If all diagnostics 038, 039, 040 are shown
in the log, this is a possible fault.
Fault: CON is not in the same state as ALA.
Description: CON is in Standby when ALA is in On (or vice versa) for more than 3.3
seconds. Technical alarm is generated.
What to do:
1.
Check if the Power switch is faulty and stuck between its fixed positions. If
the switch is sluggish, adjust the knob. If all diagnostics 038, 039, 040 are
shown in the log, this is a possible fault.
2.
Replace PC 1730 Control
3.
Replace PC 1672 Control Panel. If all diagnostics 038, 039, 040 are shown
in the log, this is a possible fault.
Fault: PAN is not in the same state as ALA.
Description: PAN is in Standby when ALA is in On (or vice versa) for more than 3.3
seconds. Technical alarm is generated.
What to do:
1.
Check if the Power switch is faulty and stuck between its fixed positions. If
the switch is sluggish, adjust the knob. If all diagnostics 038, 039, 040 are
shown in the log, this is a possible fault.
2.
Check that the PAN-PROM is correctly mounted
3.
Replace PC 1672 Control Panel. If all diagnostics 038, 039, 040 are shown
in the log, this is a possible fault.