Service procedures
MAN pressure transducer
CON
I2C-bus
Offset 1:
18. All mushroom valves except MV1 and MV5 are
opened. Also the Fresh gas valve and PEEP valve
are opened to depressurize the system.
Fail if the PEEP pressure is not within
-1 cmH2O to 0.5 cmH2O for 4 s.
Diagnostic 286, subdata ...00 02.
19. Offset adjustments of the MAN pressure
transducer E2-potentiometers until the
transducer value is close to zero.
5
The Tidal Volume display shows the target
value: 0.
The Exp. Tidal Volume display shows the
adjusting transducer value (in 1/100 cmH2O).
Fail if communication over the I2C-bus fails or
64 attempts or 30 s has elapsed to adjust
pressure transducer E2-potentiometers
without being close to the target value.
Diagnostic 285, subdata ...00 0A /...00 0B.
Gain:
20. The PEEP valve is closed to enable the system to
be pressurized.
21. The system is pressurized to +17 cmH2O. The
drive gas module regulates to a stable pressure
level with the already calibrated PEEP pressure
transducer as reference.
Fail if the PEEP pressure is not within
+16 cmH2O to +37 cmH2O for 4 seconds.
Diagnostic 286, subdata ...00 07.
22. The system pressure is increased to +20 cmH2O.
Fail if the PEEP pressure is not within
+19 cmH2O to +40 cmH2O for 4 s.
Diagnostic 286, subdata ...00 09.
5 - 14
PC1750
Transducer
Transducer
signal
23. Gain adjustments of the MAN pressure transducer
E2-potentiometers until transducer value is close
to the system pressure.
The Tidal Volume display shows the system
pressure (target value).
The Exp. Tidal Volume display shows the
adjusting transducer value (in 1/100 cmH2O)
Fail if communication over the I2C-bus fails or
64 attempts or 30 s has elapsed to adjust
pressure transducer E2-potentiometers
without being close to the target value.
Diagnostic 285, subdata ...00 0E /..00 0F.
Offset 2:
24. All mushroom valves except MV4 are opened,
also the Expiratory valve is opened to
depressurize system.
Fail if the PEEP pressure is not within
-1 cmH2O to 0.3 cmH2O for 4 s.
Diagnostic 286, subdata ...00 0D.
25. The offset adjustments of the MAN pressure
transducer E2-potentiometers is repeated, but
with a negative offset (-25 cm H
cmH2O with PC1750C) to enable measurement
of negative pressures.
The Tidal Volume display shows the target
value: 2500 (1000 with PC1750C).
The Exp. Tidal Volume display shows the
adjusting transducer value (in 1/100 cmH2O).
Fail if communication over the I2C-bus fails or
64 attempts or 30 s has elapsed to adjust
pressure transducer E2-potentiometers
without being close to the target value.
Diagnostic 285, subdata ...00 0C /...00 0D.
26. The result of the calibration of MAN pressure
transducer is shown on the KION Monitor with
the number 4 followed by Pass or Fail.
Siemens-Elema AB
KION / KION-i
O or –10
2
E382 E392E 061 01 02 02