CON: DIAG_SW_INFO
Sub data:
... AB CD EF
GH IJ
Fault: Internal program error.
Description: Debugging diagnostic log regarding software. In some cases also error
information, refer to sub data below.
What to do: Check in sub data if the diagnostic log contains any information indicating
an error. Report to EM-HSC.
Check where the diagnostic log was sent in code:
AB = Software module number
CD EF = Line number in code
GH IJ = Other information.
Error detected if:
AB
GH IJ
Description:
0C
00 0x
Failure to read A/D-port
(DRV_IN module).
28
01 00
E2-potentiometer write
failure (CAL_E2P module).
56
00 00
Hardware high pressure
detection in MEA is not
working.
54
00 00
Fresh gas zeroing cannot
be performed in manual
ventilation mode (SCON
HV module).
5A
00 00
E2-potentiometer read/
write failure at system
startup,Off to Standby
(SUP module).
Note: Technical alarm is
generated at system
startup.
What to do:
Replace PC 1730 Control.
Replace PC 1750 Transducer.
Replace PC 1748/ PC 1797 Fresh
gas flow.
Redo a PuC and check that the
Alarm detectors test works.
This diagnostic log may appear
now and then, but most of the time
the software corrects the problem
automatically by doing an extra
zeroing of the inspiratory fresh gas
flow transducer.
However, if appearing repeatedly;
1. Check that the Inspiratory fresh
gas silicone rubber valve is
correctly mounted.
2. Replace the Inspiratory fresh
gas valve with PC 1751 step
motor.
1. Replace PC 1750 Transducer.
2. Replace PC 1748/ PC 1797
Fresh gas flow.
3. Replace PC 1730 Control.