Panasonic GP-X Series User Manual page 135

High-speed high-accuracy eddy current digital displacement sensor
Hide thumbs Also See for GP-X Series:
Table of Contents

Advertisement

Ramco Innovations www.ramcoi.com
Examples of Application of RS-232C Commands
Example 1: Load the current displayed value from controller address 0.
Transmission: %EE#RMD00
Response: %EE$RMD0+000.4500
The current displayed value is "(+) 0.4500". If holding mea-
surement is designated, the hold value is sent back.
Example 2: Load the judgment data from controller address 0.
Transmission: %EE#ROT00
Response: %EE$ROT02
The current judgment is "GO".
Example 3: Load the holding measurement mode from controller address 1.
Transmission: %EE#RHM10
Response: %EE$RHM12
The current holding mode is bottom hold.
Example 4: Load the upper limit setting from controller address 0.
Transmission: %EE#RHT00
Response: %EE$RHT0+000.8000
The current upper limit setting is "(+) 0.8000".
Example 5: Enter "(+) 0.7500" as an upper limit setting for controller address 0.
Transmission: %EE#WHT0+000.7500
Response: %EE$WHT0
Example 6: Perform three-point calibration for controller address 0.
For details of three-point calibration, refer to section "4.2
Three-Point Calibration (Linearity adjustment)" on page 71.
Designate all four memories.
Transmission: %EE#WCG00
Response: %EE$WCG00
* To save only the selected memory number, send "WCG09"
instead of "WCG00" in step
After setting at zero point position, send.
Transmission: %EE#WCG01
Response: %EE$WCG01
After setting half the distance in the detection range, send.
Transmission: %EE#WCG02
Response: %EE$WCG02
CR
CR
CR
CR
CR
CR
CR
CR
CR
CR
CR
.
CR
CR
CR
CR