4.5 Analog Voltage Output Scale

The full-scale is set in the range from 0 to 5.0V before shipment from the fac-
tory.
An analog voltage output range, inclination and other features of the detect-
ed value can be entered.
The setting method includes three variations, similar to the display scale:
one-point scale, two-point scale, and INV.
Or select the default setting to return to the factory shipment state.

4.5.1 One-Point Scaling

unchanged. With one-point scaling, a "SLOPE" function is available with
which the inclination reverses around the point designated for scaling.
Change "0.5mm 1.2500" to "0.5mm 1.5000".
Voltage
(V)
5.0000
1.2500
0
0.5mm
Ramco Innovations www.ramcoi.com
Only the voltage output can be changed with this function. To change the
displayed value according to the detected value, use the display scale.
Notice
However, if voltage (volts) is selected as a display unit, the indication
changes after analog voltage output scaling.
2mm
Voltage
(V)
5.2500
5.0000
1.5000
0.2500
0
0.5mm
2mm
Phone 800-280-6933 nsales@ramcoi.com
Change the inclination with "SLOPE".
Voltage
(V)
5.2500
1.5000
0.2500
-2.2500
0.5mm
2mm
81