High-speed high-accuracy eddy current digital displacement sensor
Hide thumbs Also See for GP-X Series:
Table of Contents

Advertisement

*To perform zero setting, set the detection distance within 1/2 F.S. If the de-
tection distance between the measurement result and zero setting exceeds
1.5 times F.S., calculation may not be performed correctly.
<Zero setting without display scale>
(Example) GP-XC10M with 2mm detection distance
Displayed value
2.0
0
<Zero setting with display scale offset at detection distance 0mm>
Displayed value
2.0
0
Note: In the case of zero setting at the external terminal, zero setting is not
canceled but performed again.
Similar change occurs in the case of analog voltage output scale.
Ramco Innovations www.ramcoi.com
The memory can be overwritten 100,000 times. Be careful that it will not
be overwritten too frequently by the zero setting input from an external ter-
Notice
minal. Furthermore, it is recommended that the panel key lock function be
turned ON to prevent inadvertent overwriting.
Press and hold
key at 1.0mm.
Press
Detection
key again to
distance
restore to the
original state.
2.0
(Note)
Changed arbitrarily
according to scale.
Refer to section
Detection
"4.4 Display Scale"
distance
on page 76 and
section
2.0
Analog
Voltage
Output Scale" on
page 81.
Offset value
Displayed value
2.0
and
0
1.0
2.0
After zero setting
Displayed value
2.0
Detection
distance
0
"4.5
2.0
1.0
3.0
Example after display scaling
Phone 800-280-6933 nsales@ramcoi.com
Detection
distance
Displayed value
Press and hold
key at 1.0mm.
Press and hold
key again
0
to restore the
original setting.
(Note)
Due to scaling,
the set detection
distance
0mm
becomes
the
datum for zero
setting.
After zero setting
Detection
distance
1.0
2.0
29