Probes and Biopsies
5.2.5 Biopsy guide introduction
Biopsy
PEC63
H48681GF
PEC71
PEC74
5-18
6. Select the [Store MBX-3] key to store the second line.
7. Exit to the 2D main menu without saving.
8. Select the [Exit + Save] key to store biopsy setup.
The MBX-2 line will be automatically calculated.
It is absolutely necessary to ensure that before performing a biopsy, the selected and
displayed biopsy line corresponds to the biopsy needle guide mounted to the transducer (left/
right).
If the material of the biopsy needle guide is plastic, only Single-Use is possible!
The biopsy bracket is reusable.
Please refer to the manufacturer's instructions included in the biopsy kit.
Probe
Capabilities and Features
RIC5-9W-
Needle diameters: < 1.8 mm
RS
Material: Stainless Steel
Sterilization with autoclave possible!
RIC5-9W-
Needle diameters: < 1.65 mm
RS
Material: Plastic
Sterile packaged component. Single-Use
only!
AB2-7-RS
Needle diameters: < 1 mm < 1.4 mm <
2.2 mm
Material: Stainless Steel
Sterilization with autoclave possible!
RAB4-8-RS
Needle diameters: < 1 mm < 1.4 mm <
2.2 mm
RAB2-5-RS
Material: Stainless Steel
Sterilization with autoclave possible!
Illustration
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4