Utilities and System Setup
Services:
Alias:
AE-Title:
IP-Address:
Port Number:
Storage Commit:
Send Sequ.:
13-34
Select [STORE] to send screen images, 2D cine sequences and 3D/4D
data to a DICOM server (e.g., Radworks Server).
Select [STORE 3D] to send 3D/4D data only (volumes and cine
sequences) to a different store server (e.g., PC with Software 4D View
installed) than screen images and 2D cine sequences.
Select [PRINT] to send images stored in printer clipboard to a DICOM
printer.
Select [MPPS] to send images to a DICOM server with transfer
information.
Select [ST. COMMIT] to send image with an additional layer of security.
Select [STR. REPORT] to send the Patient report data to a PC via
network or serial port.
Select [WORKLIST] to retrieve Patient Information (Name, ID, Birth,...)
from an external Worklist server (e.g., HIS - Hospital Information
System / RIS).
Select [VIEWPOINT] to get default settings for Viewpoint server.
Enter a nickname for each DICOM node to make it easier to handle
various nodes. Use any name, but do not insert space characters.
Application Entity Title of the remote DICOM application.
Enter the hostname or IP-address of the DICOM node. e.g.,
any.dicom.server.net
Enter the port number of the DICOM node. (e.g., 104)
The "Storage Commit" drop down menu contains all currently available
storage commit servers. The selected storage commit server is used for
committing the images sent to this store server.
If "Send sequ." is checked, it causes all data to be sent to this server
sequentially. This means that only one transfer is active at a time. If one
transfer fails, all subsequent transfers are stopped until the failed transfer
succeeds or is removed from the queue. (Use for servers that cannot
handle multiple associations, or do not sort the images by Image Number.
2If "Send sequ." is not checked, the up to 5 data sets can be transferred
at the same time. This means that transfer is faster. Images can arrive
out of order in this case. (Use for servers that have none of the limitations
listed in the above paragraph).